Inhalation April 2026 17 air mass flow rate in standard liters per minute (stan- dard conditions T = 21.1 ÂșC, P = 101.3 kPa). e mass flow rate through the Parker valve 8FV12LN is indeed choked when the ratio of the downstream pressure to the upstream pressure is smaller than approximately 0.5. A more complete academic anal- ysis would quantify and display the slope of the curve of mass flow rate versus pressure ratio, but the "eye- ball" reaction to the data in Figure 3 is sufficient for the current discussion. ere are two curves in Figure 3. e round data points show the results when the Parker valve is in the controlled test bed shown in Figure 2. e triangular data points show the results where the same model of Parker valve is the flow control valve installed in a commercially available flow control unit. e curves do not overlap in Figure 3 only because the valve under test was more open when tested in the com- mercial flow controller equipment than it was when tested alone in the designed test bed. e import- ant matters are that there is a change in slope from approximately parabolic (to the right of 0.5) to flat (to the left of 0.5) and that this change takes place for both data sets at a pressure ratio of approximately 0.5. (e theory for air is noted at a pressure ratio of 0.528, as given by equation 7, below.) e reason to show both curves here is not to call out a particular commercial product but only to high- light that the same method described for the test bed also serves to test commercial flow control equip- ment. e test of any similar valve in a designed test bed (Figure 2) and the test of that valve in its com- pressure transducer (Mensor LP, San Marcos, Texas, US) (item 5). e downstream absolute pressure, P 2 , is then calculated from the ambient pressure minus the differential pressure reading. e test protocol calls for the valve under test to be open a fixed amount, the vacuum pump to be turned on, and the flow rate adjusted with the system con- trol valve so the downstream pressure is only about 30% of the ambient pressure. Achieving these con- ditions requires some trial and error when testing a valve with unknown behavior. Once the desired range of operation is set, the valve under test must remain unchanged. Now, the first data point with the down- stream pressure at about 30% of the upstream pres- sure is measured by recording the mass flow rate of air entering the system along with the pressure upstream and downstream of the valve under test. Next, the sys- tem control valve is closed slightly until the pressure downstream of the valve under test increases to 35% of the upstream pressure, then to 40%, and continues in increments of 5% until the downstream pressure is approximately 70% of the upstream pressure. At each value of the downstream pressure, the user records the mass flow rate of air entering the system and the pressure upstream and downstream of the valve under test. In this way, the pressure ratio across the valve under test increases from about 0.3 to about 0.7 and, importantly, thoroughly covers values in the range of the ratio 0.5, where theoretically the flow changes from choked to unchoked. Figure 3 shows data with this apparatus when testing the Parker valve 8FV12LN. Here, the y-axis is the Figure 3 Onset of choked flow for Parker Valve 8FV12LN installed in commercial equipment and in a stand-alone valve test bed Mass Flow Rate, Standard L/min 62.0 60.0 58.0 56.0 54.0 52.0 50.0 48.0 46.0 0.20 0.30 0.40 0.50 0.60 0.70 0.80 Pressure Ratio, P d /P u Installed Test Bed Theory for Air 0.528